Key terms for this module
Read these first. Every one of them comes up in the sections below.
| UV sterilizer | A unit that passes water past a UV-C lamp. |
| UV-C | Short-wave ultraviolet light that damages DNA. |
| Quartz sleeve | The clear tube that keeps water off the UV lamp. |
| Dose | The amount of UV the water gets, in µWs/cm². |
| Ballast | The electrical part that powers the UV lamp. |
| Protozoa | Single-celled organisms, including many parasites. |
| Phytoplankton | Tiny free-floating algae; food for filter feeders. |
| Bacterioplankton | Bacteria living free in the water column. |
| Clarifier | A UV run at higher flow mainly to clear green water. |
| Bucket test | Timing how long a line takes to fill a known volume. |
| Automatic water changer | A unit that drains and refills water on a schedule. |
| Siphon break | A part that stops water flowing back by gravity. |
| Automatic feeder | A unit that drops a set portion of food on a schedule. |
| Cold food feeder | A fridge and pump that feed liquid or frozen food on a schedule. |
1. What UV sterilization does
A UV sterilizer passes water past an ultraviolet lamp that damages the DNA of what floats through.
A UV sterilizer pumps water past a lamp that gives off UV-C light, inside a quartz sleeve that keeps the water off the lamp. UV-C damages the DNA of organisms suspended in the water passing through, so they can't reproduce. It reduces free-floating populations; it doesn't wipe them out, and it removes nothing from the water.
Only what passes the lamp is affected. Bacteria living on rock, sand and filter media stay on those surfaces and aren't exposed, and neither are fish, rock or glass. Everything drifting in the water is exposed, including plankton and the spawn of invertebrates (section 6). Like low-micron filtration (Module 15), UV can only reach the free-swimming stage of a parasite, never the stage on the fish. Module 16 covers ozone, the other oxidizing treatment, and Module 11 covers which materials stand up to UV.
| If | Then |
|---|---|
| Parasites on the fish | UV can't reach them there. |
| Worried UV kills the biofilter | Bacteria on surfaces aren't exposed. |
Check your understanding
A fish has ich spots. Will turning on a UV cure it?
No. UV only reaches the free-swimming stage that passes the lamp, not the parasites on the fish. Treatment is Module 30.
Key point Only what passes the lamp.
2. Dose and flow
The slower water passes the lamp, the higher the dose it gets.
UV dose is measured in microwatt-seconds per square centimeter (µWs/cm²). Free-floating algae and bacteria need a lower dose, about 30,000 µWs/cm²; protozoa, which include parasites, need about 180,000 µWs/cm². The dose depends on how long water spends in front of the lamp, so slower flow means a higher dose.
Makers give two flow rates: a faster one for algae and bacteria and a much slower one for parasites. For example, a 25 W unit lists about 1,790 L/h (472 US gallons per hour) for algae and bacteria, but about 300 L/h (79 US gallons per hour) for protozoa. Measure the flow, don't guess: time how long the line takes to fill a 19 L (5 US gallon) bucket. One minute is about 1,140 L/h (300 US gallons per hour); two minutes is about 570 L/h.
| If | Then |
|---|---|
| Flow unknown | Bucket test. |
| Targeting parasites | Much slower flow. |
Check your understanding
A UV is set up to fight parasites, but the line fills a 19 L bucket in 30 seconds. What's wrong?
That's about 2,300 L/h, far faster than a parasite flow rate. Slow it to the maker's protozoa rate with a valve.
Key point Slower flow, higher dose.
3. Sizing a UV
Choose the target, then a unit big enough for the system, then set its flow.
Decide what the UV is for: clear green water and bacteria, or reduce parasites. Higher wattage means more UV and a bigger unit. Never choose a unit whose maximum system size is smaller than your real water volume (Module 36); a larger one is fine. Then set the flow to the maker's rate for your target, with a pump that can reach it against its plumbing and a valve to set it.
One unit can't do both targets at once, because each needs a different flow. Either pick one, or run two units at different flows. Some makers deliver a stronger dose at their listed flows than others, so compare the published flow and dose, not just the watts.
| If | Then |
|---|---|
| Unit rated below system volume | Too small. |
| Want algae and parasites | Two units or two flows. |
Check your understanding
A UV is rated for systems up to 250 L. The system's real volume is 400 L. What's the problem?
It's too small. Choose a unit rated for at least the real water volume.
Key point Target, then size, then flow.
4. Plumbing and placement
Plumb a UV for the right flow, keep it full of water, and feed it water it hasn't just treated.
The most common setup is inline on the return pump line: simple, and hidden in the cabinet, but the flow is whatever the return gives. A dedicated pump with a valve, or a controllable DC pump, sets the flow exactly. A closed loop drawing straight from the display is the most effective but least convenient.
Draw water from one sump chamber and return it downstream, such as from the first chamber to the return chamber, so the UV isn't re-treating the same water. Mounted horizontally, the inlet and outlet face up; mounted vertically, water enters at the bottom and leaves at the top. Trapped air heats the unit and cuts its effect. Secure the lines so they can't fall out of the sump, and never switch the lamp on until water is flowing through it.
| If | Then |
|---|---|
| Inlet and outlet in one chamber | Re-treating the same water. |
| Lamp on, no flow | Switch off: it overheats. |
Check your understanding
A UV draws from the return chamber and returns to the return chamber. What happens?
It re-treats the same water. Draw from an earlier chamber and return downstream.
Key point Full of water, right flow.
5. When to run UV
UV can run all the time, on a timer, or only when there's a problem.
A UV can run all day, run on a timer for part of the day, or sit ready and run only when a problem appears. Common reasons to run one are new animals being added, green water, a disease outbreak and dinoflagellates. It reduces what's floating; it doesn't cure fish that are already sick (treatment is Module 30).
Whether to run it all the time is a separate decision, covered in the next section. Whatever the schedule, switching a UV on or off is a change to a running system: make one change at a time and watch the result (Module 1).
| If | Then |
|---|---|
| Adding new animals | Run it to lower disease risk. |
| Sick fish | UV doesn't cure them (Module 30). |
Check your understanding
New fish are being added next week. Is that a reason to run the UV?
Yes. Running UV while stocking lowers the risk of disease spreading, though it won't cure a fish that's already sick.
Key point All day, timer or as needed.
6. Running UV all the time
Running UV all day keeps free-floating problems down, but it also kills free-floating life the system needs.
Reasons to run it all the time. It keeps free-floating algae, bacteria and free-swimming parasites down every hour of the day, not only after a problem shows. That matters most while new animals are being added, and in quarantine systems. It also keeps the water clear and can help against dinoflagellates.
Reasons not to. UV doesn't choose what it kills. Phytoplankton and bacterioplankton that pass the lamp die along with the nuisance species. They're a natural food for many filter feeders, corals included, and while alive they take up nitrate and phosphate, so with fewer of them nutrients can climb. Most aquariums already hold little plankton, and a UV running all day keeps it lower still. That's the cost to biodiversity: less free-floating life, and less food for the animals that depend on it.
How much that costs depends on the system. A system started with dry rock holds little invertebrate life and little plankton, so there's little for UV to take. A high-biodiversity system is built the other way: started with uncured live rock taken straight from the ocean, and fed a wide range of foods, from bacteria through zooplankton. Many hold living plankton, including spawn from broadcast-spawning invertebrates. A UV in a system like that will dampen the biodiversity the system is built on.
Switch the UV off for these:
- Medication: UV can break medications down, chelated copper in particular. Switch it off for the treatment and follow the medication's own directions (Module 30).
- Dosed bacteria: bacteria products go into the water column, where the UV reaches them. Some makers direct switching off the UV, often with the skimmer, for 12 to 48 hours after dosing, depending on the product. Follow the label.
- Live phytoplankton and rotifers: food dosed for filter feeders passes the lamp too. Feed while the UV is off, so the food reaches the animals first.
- A new system that's still cycling: UV can kill nitrifying bacteria in the water before they settle on rock and media (Module 9).
A timer is the middle ground: the UV runs part of the day and stays off while food, bacteria or medication is in the water. Log every switch-off with its planned switch-on time (Module 20), so a UV is never left off by mistake. Nuisance algae growing on rock and glass never pass the lamp, so UV doesn't replace good maintenance.
| If | Then |
|---|---|
| Medication in the water | UV off; follow its directions. |
| Dosing bacteria | UV off for the label's time. |
| High-biodiversity system | UV dampens it. |
Check your understanding
A system is being treated with chelated copper, and the UV runs all day. What's wrong?
UV can break the medication down. Switch the UV off for the treatment and follow the medication's directions.
Corals are fed live phytoplankton every morning, and the UV runs all day. What would you change?
The phytoplankton passes the lamp. Put the UV on a timer that keeps it off while feeding, and log when it comes back on.
Key point UV doesn't choose.
7. UV maintenance
The lamp and its quartz sleeve both age. Replace them about once a year.
UV lamps lose output with age, and the quartz sleeve gets a film that blocks UV-C. Replace the lamp and the sleeve about once a year, following the maker. Green water coming back after months of clear water is a sign they're due. Clean the sleeve whenever the unit is open.
Safety: unplug the unit before opening it. Handle the lamp with gloves or a towel so skin oil doesn't mark the glass. UV-C harms eyes and skin; check that the lamp is on through the unit's viewing window, never by looking at a bare lamp. After reassembling, check the cap and seals for drips, let water fill the unit, then switch it on.
| If | Then |
|---|---|
| Lamp over a year old | Replace lamp and sleeve. |
| Checking the glow | Window only, never a bare lamp. |
Check your understanding
A coworker opens a running UV to look at the lamp. What's wrong?
Unplug it first, and never look at a bare UV-C lamp. Check the glow through the viewing window.
Key point Unplug first.
8. UV problems
When a UV stops working, check power and lamp, then flow, then the sleeve and the plumbing.
No glow in the window: check the outlet and controller (Module 20), then the ballast and lamp. Green water coming back after months of clear water: the lamp and sleeve are aging or the sleeve is filmed. No effect at all: the flow may be far too fast for the target, or the inlet and outlet are in the same chamber so the same water goes round and round. Unit hot or noisy: trapped air or no flow; switch it off until water is flowing. A drip at the cap: the O-rings or seals need reseating or replacing.
Measure the flow with the bucket test before and after any change, and change one thing at a time. A UV can sit dark for weeks without anyone noticing, so its glow is part of the daily equipment check (Module 20).
UV drill: on the course website this is an interactive drill. Read the symptom, check the lamp, sleeve, flow and plumbing, and fix the cause.
| If | Then |
|---|---|
| No glow | Power, ballast, lamp. |
| Drip at the cap | O-rings and seals. |
Check your understanding
The water was clear for a year, and now it's turning green again. The UV glows. What do you check?
The lamp's and sleeve's age and the sleeve for film. Replace them about once a year.
Key point Power, flow, sleeve, plumbing.
9. Automatic water changers
An automatic water changer drains a set amount and refills it with new salt water.
An automatic water changer (AWC) uses level sensors and pumps. A typical unit has two optical sensors in the chamber where the level moves (usually the return chamber): it pumps water out to a waste container or drain until the low sensor trips, then refills from a container of new salt water until the high sensor trips. Some swap water in and out at the same time instead. Many combine with the top-off, using the high sensor to add fresh water between changes. Siphon breaks and secure line holders stop water flowing back by itself.
The new water must match the system's salinity, temperature and alkalinity, just like a manual water change. The big risks are the wrong water in the new-water container, a container that runs empty, and a waste line that siphons. The top-off failsafes in Module 20 apply here too: check the containers and sensors in the daily equipment check, and watch one full change after any setup work.
| If | Then |
|---|---|
| New water container empty | Refill before the next change. |
| New water unmatched | Match salinity, temperature, alkalinity. |
Check your understanding
After an automatic water change, salinity jumped 2 ppt. What went wrong?
The new water didn't match. Check the new-water container's salinity before the next change.
Key point Drain to low, refill to high.
10. Automatic feeders
An automatic feeder delivers a set amount of food on a schedule. It doesn't watch the fish eat.
Automatic feeders hold dry food and drop a set portion on a schedule. Some keep food dry with a closed drawer, use different tray sizes to set the amount, can pause the pumps during feeding, and can warn when food runs low. Food that gets damp clumps and stops dropping, so check that each feeding actually dropped.
Dry food is only part of a varied diet. Cold food feeders keep liquid or thawed frozen food in a fridge and pump a set amount into the sump on a schedule, with a peristaltic pump like a dosing pump (Module 20). The food is kept as close to freezing as possible without freezing, because unpreserved food goes bad in less than a week. Wide pump tubing and finely ground food of even size clog less. Tank water must never run back up the line into the food: it contaminates it. These feeders make it easy to overfeed, so watch nitrate and phosphate, and keep flow and mechanical filtration strong.
Feeding is also a health check. A healthy clownfish springs out and eats when food goes in, even if it was resting on the bottom. A feeder can't see that, so it adds to hand feeding rather than replacing it; someone still watches the animals eat and removes food left uneaten. Live and frozen foods are covered in Module 26.
| If | Then |
|---|---|
| Food clumped in the feeder | Damp: clean, dry, refill. |
| Fish not eating | Health check. Find out why. |
| Cold feeder line clogged | Grind finer; wider tubing. |
Check your understanding
The feeder ran on schedule, but the clownfish didn't come out to eat. What do you do?
Treat it as a health warning: check the fish and the water. A healthy clownfish springs out to eat.
A cold food feeder was added a month ago, and nitrate and phosphate have climbed since. What do you check first?
The amount it feeds. Cold food feeders make it easy to overfeed; cut the portion and check that flow and mechanical filtration are strong.
Key point Watch them eat.
11. What you see, and what it means
| No glow in the UV window | Power, ballast or lamp. |
| Green water back after a year | Lamp and sleeve due. |
| UV hot, no flow | Switch off; trapped air or no flow. |
| Drip at the UV cap | O-rings or seals. |
| Medication in the water | UV off; it can break it down. |
| Corals fed live phytoplankton | UV off while feeding. |
| Parasites despite UV | Flow too fast, or stage on the fish. |
| Salinity jump after AWC | New water unmatched. |
| AWC didn't refill | Empty container or sensor. |
| Feeder dropped nothing | Damp, clumped food. |
| Fish didn't come to eat | Health warning. |
| Nitrate up after a cold feeder | Overfeeding. |
Module 21 test
30 questions. Pass with 27 or more. Opens after you pass every section check in this module. Sign in to start.
Sources
- Bulk Reef Supply: A guide to choosing the right UV sterilizer
- Bulk Reef Supply: Adding a UV sterilizer to your reef tank can be easy
- Aqua Ultraviolet 40 W Classic: flow, lamp and sleeve replacement (Bulk Reef Supply listing)
- AutoAqua AWC auto water changer plus ATO (Bulk Reef Supply listing)
- GHL: Automatic Fish Feeder
- AlgaeBarn: Should I use ultraviolet sterilizers?
- LiveAquaria: UV sterilization selection guide
- Evolution Aqua: Do I need to turn the UV off when using medications?
- DrTim's Aquatics: What to use and not use together
- Reef2Reef: Automatic frozen and refrigerated food dispenser
- Michael Maddox, President of IMARCS: clownfish feeding behavior, matched water changes, daily equipment checks, high-biodiversity systems and UV