Module 7

Testing and Measurement

How numbers go wrong, accuracy and precision, detection limits, sampling, test kits, photometers, calibration, salinity tools, probes, automated testers, lab tests, confirming strange numbers, testing routines and trends.

Free to read. No account needed.

Sign in

Key terms for this module

Read these first. Every one of them comes up in the sections below.

Accuracy How close a result is to the true value.
Precision How closely repeated tests of the same water agree.
Resolution The smallest step a test shows, like 0.1 dKH.
Detection limit The lowest amount a test can tell apart from zero.
Reagent The chemical in a test that reacts with the water.
Titration Adding reagent bit by bit until a color change shows the result.
Endpoint The color change that ends a titration.
Photometer A meter that reads a test by how much light gets through the vial. Also called a colorimeter or checker.
Cuvette The glass vial, or the slot it sits in, in a photometer.
Calibrate Set a tool against a solution of known value.
Verify Check a tool by reading a fresh solution of known value.
Standard A solution made to an exact known value. For pH, called a buffer.
Probe The part of a meter or controller that sits in the water.
Offset How far a tool reads off, written on its label. Apply it before recording.
Refractometer A tool that reads salinity from how much water bends light.
ICP A lab test that measures many elements at once from one sample.
Drift A number creeping slowly in one direction over weeks.
Graduated cup A cup marked with lines for measuring volume.

1. Every number can be wrong

A test result is only useful if it's true. A wrong number is worse than no number, because people act on it.

Where a number can go wrong The sample Wrong water, wrong amount, or left sitting too long The tool Dirty, expired, not calibrated The reading Read too early, misread, wrong unit The record Wrong tank, wrong box, typo Four places a number can go wrong. Most strange numbers are test errors.

A number can go wrong in four places: the sample (the wrong water, the wrong amount, or water left sitting), the tool (dirty, expired or not calibrated), the reading (read too early, misread, or in the wrong unit) and the record (the wrong tank, the wrong box, or a typo).

When a number looks strange, don't ask "what's wrong with the tank?" first. Ask "what could be wrong with the number?" Check the sample, the tool, the reading and the record. Most strange numbers are test errors.

If Then
A number looks strange Check the sample, the tool, the reading and the record first.
You assume the strange number is the tank You may "fix" a tank that was fine.

Check your understanding

Nitrate was 12 ppm last week. Today it reads 40 ppm. Nothing about the tank looks different. What do you check before you believe it?

The sample, the tool, the reading and the record. A jump that big with no other change is most likely a test error.

Key point Before you ask what's wrong with the tank, ask what's wrong with the number.

2. Accuracy, precision and resolution

Three different words describe how good a test is. Mixing them up leads to trusting numbers you shouldn't.

Accuracy and precision Accurate and precise Precise, not accurate Accurate on average, not precise Neither Center: the true value. Dots: repeated tests. Precise repeats itself. Accurate hits the truth.

Accuracy is how close a result is to the true value. Precision is how closely repeated tests of the same water agree with each other. A test can be precise and still inaccurate: it gives nearly the same wrong number every time. Resolution is the smallest step a test shows, like 0.1 dKH or 0.01 ppm. A display showing 0.01 steps isn't promising the result is right to 0.01.

One reading is a range Alkalinity checker: one reading of 8.0 dKH 6 7 8 9 10 8.0 Shaded: 7.3 to 8.7, where the true value could be (stated accuracy: 0.3 dKH plus 5% of the reading) ! A range, not exact Treat one reading as a range. Compare with earlier readings from the same tool, and retest big moves.

Makers state accuracy as a range. One marine alkalinity checker states its accuracy as 0.3 dKH plus 5% of the reading. So when it reads 8.0 dKH, the true value could be anywhere from about 7.3 to 8.7. One reading is a range, not an exact number.

That's why a single reading isn't acted on alone. Compare it with earlier readings from the same tool and method, and retest any big move before writing it down (Module 6).

If Then
Repeated tests agree closely The test is precise. That doesn't prove it's accurate.
The display shows 0.01 steps That's resolution, not accuracy.
One reading of 8.0 dKH The true value is a range around it. Compare with earlier readings.

Check your understanding

A checker gives 8.1, 8.1 and 8.0 dKH on the same water. A fresh standard of 8.0 reads 9.0. Is the checker precise? Accurate?

Precise: it repeats itself. Not accurate: it reads about 1 dKH high. It needs checking and calibrating before it's trusted.

Key point Precise repeats itself. Accurate hits the truth.

3. Detection limits: what zero means

Every test has a point below which it can't tell a small amount from none at all. Below that point, a result of zero doesn't mean zero.

Detection limit Phosphate checker, low range (ppm) 0.00 0.02 0.04 0.06 0.08 0.10 Shaded: too close to zero to tell apart from zero on this test ! 0.00 isn't "none" It means below what this test can see.

The detection limit is the lowest amount a test can reliably tell apart from zero. One low-range phosphate checker shows results to 0.01 ppm, with a stated accuracy of 0.02 ppm plus 5% of the reading. A reading of 0.00 on that test means "less than about 0.02", not "no phosphate".

This matters because zero nitrate or phosphate isn't normal (Module 6). A reading at or below the detection limit is reported as low, the same as a reading of zero. Some lab reports print 0.00 for anything below their limit; "not detected" would be the more honest label.

If Then
Phosphate reads 0.00 Below what the test can see, not "none". Report it as low.
A lab report shows 0.00 Likely below its detection limit.

Check your understanding

Phosphate reads 0.00 ppm on a low-range checker. A coworker logs "no phosphate, perfect." What's wrong?

0.00 means below what the test can see, not none. And zero isn't normal: it's reported as low.

Key point Zero on a test means "too low to see".

4. Taking a good sample

A perfect test on the wrong water gives a perfect wrong answer. The sample has to represent the system.

Taking a sample A good sample Mixed water from the main tank, clean container, tested fresh Same time of day Match the readings you compare. ! Not at an outlet Dosing or top-off water isn't mixed ! Not left sitting It can change. Test it fresh.

Take the sample from well-mixed water in the main tank, in a clean container, and test it fresh. Water right at a dosing line tip or a top-off outlet hasn't mixed with the rest of the system yet, so it reads higher or lower than the tank really is. Water left sitting before testing can change, so test right away.

Take it at about the same time of day as the readings you'll compare it with. pH and some other readings change through the day (Module 6).

If Then
Sample taken at the dosing tip Not mixed water. Take it from the main tank.
Sample left sitting an hour It can change. Test fresh.

Check your understanding

A sample is taken right where the alkalinity dosing line drips into the sump. The result is high. What happened?

That water hasn't mixed with the system. Take the sample from mixed water in the main tank and retest.

Key point Mixed water, clean container, tested fresh.

5. Test kits: titration and color matching

Many hobby test kits work by adding reagent to a sample and reading a color. How they're read matters as much as the chemistry.

How a titration kit works A measured sample Add reagent bit by bit The color changes This is the endpoint. The result From how much reagent was used ! Color judgment varies Same kit, white background, room light, not tank light

A titration kit adds a reagent of known strength to a measured sample, a little at a time, until the color changes. That color change is the endpoint. How much reagent it took gives the result. Alkalinity kits commonly work this way. Other kits add a fixed amount of reagent and match the color to a printed chart.

Both depend on people judging color, and people see color differently, especially in poor or colored light. Read the color against a white background under ordinary white room light, not under the blue aquarium lights. In one side-by-side comparison of alkalinity kits on the same water, results ranged from about 10% low to about 30% high, depending on the brand. So compare readings from the same kit, and if you change kits, test with both for a while to see how they compare.

Never pour unused reagent back into its bottle: it can contaminate the whole bottle. Reagent loses strength with age, so check the date.

If Then
Two brands give different results Kits differ. Compare readings from the same kit.
Leftover reagent Never pour it back into the bottle.
Reading color under the tank lights Colors look different. Read under white room light.

Check your understanding

The facility switches to a new brand of alkalinity kit. The first reading is 0.8 dKH lower than last week. Is alkalinity falling?

Not necessarily. Different kits read differently on the same water. Test with both kits on the same water to see how they compare before deciding.

Key point Same kit, white room light, fresh reagent.

6. Photometers

A photometer, also called a colorimeter or checker, reads the test with light instead of your eyes. It removes color judgment, but it adds its own ways to go wrong.

How a photometer reads Light source Sample vial Detector The meter measures how much light gets through the vial. ! Keep the light path clear Prints, spots, bubbles, scratches, stains and outside light all change the number.

The meter shines light through the vial of reacted sample and measures how much gets through. Anything else in the light path changes the number: fingerprints, water spots, bubbles, scratches, stains, or outside light getting in.

Hold vials by the top and wipe the outside clean and dry before each reading. Tap off bubbles. Put the vial in the same way every time, close the lid fully, and keep the meter's well free of dust. After each test, empty and rinse the vial; reagent left sitting in a vial can stain the glass for good. When a reading is unexpected, check the meter with the maker's calibration set of known values.

If Then
Spots, prints or bubbles on the vial Light is blocked. Clean, dry and retest.
Lid left open Outside light changes the reading.
Reagent left in the vial for hours It can stain the glass for good.

Check your understanding

A phosphate test gives 0.45 ppm. Last week was 0.05. The vial has white water spots on it. What do you do?

The spots block light. Rinse, wipe the vial clean and dry, and test again. If it's still far off, check the meter with its calibration set before recording.

Key point Clean vial, same position, lid closed.

7. Running a test the same way every time

A clean meter and a clean vial still give a wrong number if the test isn't run the same way every time. Each test has a method: follow it exactly.

Running a test Right test chosen on the meter Sample exactly to the line All of the reagent, fully mixed The full reaction time Reagent within its date Same way, every time. Skip one and the number still looks real.

Most test errors come from the same few things: the wrong test chosen on the meter, the wrong sample size, reagent not fully added or mixed, reading before the reaction time is up, or expired reagent. Each one gives a number that looks real but isn't.

FOLLOW THE METHOD. Sample to the line, all of the reagent, the full reaction time, every time. A retest costs a few minutes. A wrong number in the log costs far more.
If Then
Sample filled past the line Wrong sample size, wrong result.
Read before the reaction time is up The color may still be changing.
Reagent past its date A wrong number that looks normal.

Check your understanding

Two staff test the same tank's alkalinity ten minutes apart and get 7.9 and 6.8 dKH. Nothing changed in the tank. What happened?

The tank can't change that fast, so at least one test is wrong: sample size, reagent, reaction time or a dirty vial. Retest carefully and confirm before recording.

Key point Same method, every time.

8. Measuring liquids: the standard cup

Doses, reagents and small amounts of water all have to be measured. A wrong measurement is an error just like a wrong reading.

The standard measuring cup The standard 50 mL measuring cup One cup with two measuring areas 10 mL 20 mL 30 mL 40 mL 50 mL 1 2 3 4 5 Rear area: 1 mL lines Main area: 5 mL lines to 50 mL Shape only, not to scale. 5 mL or less Use the small rear area Over 5 mL Use the main area lines Read it right Flat surface, eye level, at the line

The standard measuring device for liquids is a 50 mL graduated cup with a separate 1 mL section, such as the one that comes with Red Sea Reef Energy AB+. It's one cup with two measuring areas: the main area, marked in 5 mL steps up to 50 mL, and a small compartment at the back, marked in 1 mL steps up to 5 mL.

Use the small rear compartment for amounts of 5 mL or less, and the main area for anything larger. Small amounts can't be read accurately on the big lines. Set the cup on a flat surface and read it at eye level, with the liquid at the line. Use a clean cup, and rinse it in RO water after use, so one liquid doesn't carry into the next.

Use the same standard cup every time, so a measured amount means the same thing whoever measures it. A dose measured in a different cup or by guessing can't be compared or repeated.

If Then
Measuring 3 mL Use the small rear compartment, not the main area.
Measuring 30 mL Use the main area lines.
Reading the cup held up in the air Set it on a flat surface and read at eye level.

Check your understanding

A dose of 4 mL is measured by pouring to just under the first line of the main cup. What's wrong?

Small amounts can't be read accurately on the main cup lines. Use the small rear compartment, marked in 1 mL steps up to 5 mL.

Key point Rear compartment up to 5 mL, main area above. Flat, eye level, clean cup.

9. Calibration and verification

A tool can be clean, switched on and showing a number, and still be wrong. The only way to know it's right is to test it against something whose value is already known.

Calibrate, then verify Calibrate Set the tool against a known value. Verify Read a fresh known value. Matches Trust the tool. ! Doesn't match Clean, recalibrate, verify again. Calibration without verification is a guess.

Calibrating means setting a tool against a solution of known value, so it learns what that value looks like. Verifying means checking the tool afterwards, by reading a fresh solution of known value and seeing if the number matches.

Calibration solutions, called standards (or buffers for pH), are made to an exact value. They only work fresh. Pour out only what you need into a clean container, use it once, and never pour it back into the bottle.

pH probes are calibrated with at least two buffers, usually one to three pH units apart, for example 7.00 and 10.01. Rinse the probe in RO water between buffers so one doesn't carry into the next. Some meters show a slope after calibrating; 92% to 102% is acceptable. Outside that range, don't trust the probe: report it.

Calibrate and verify on the schedule your facility or the maker sets, and any time a reading can't be trusted. A second device checked against a standard can also verify a reading: if two trusted devices agree, the number is real.

If Then
Calibrated but never verified Nobody knows if the calibration worked.
Used standard poured back The whole bottle is contaminated.
pH slope outside 92% to 102% Don't trust the probe. Report it.

Check your understanding

You calibrate a salinity meter, then read a fresh 35 ppt standard and get 32.8. What now?

The calibration didn't take, or the standard used was bad. Recalibrate with fresh standard and verify again. Measure nothing until it reads close to 35.

Key point Calibrate against a known value, then verify with a fresh one.

10. Measuring salinity

Salinity is measured with a conductivity meter or a refractometer. Both need calibrating at the salinity you actually measure.

Measuring salinity Conductivity meter Refractometer Calibrate 35 ppt standard Calibrate 35 ppt fluid Bubbles Tap them off the probe. Temperature Let it settle first. Rinse RO water after use Rinse Prism with RO water Either way: calibrate near the salinity you measure, then verify.

A conductivity meter reads salinity from how well the water carries electricity. Calibrate it with a 35 ppt standard: put the probe fully in, tap off bubbles and let it finish. Rinse it in RO water before measuring anything.

A refractometer reads salinity from how much the water bends light through a prism. Calibrating it with 35 ppt calibration fluid is generally preferred over RO water, because refractometer scales aren't perfectly even across their range, so setting it near reef salinity is more accurate there. Use RO water for rinsing the prism. Give the sample about 30 seconds on the prism to reach the prism's temperature, and don't read right after carrying the refractometer between a cold room and a hot one. Models with ATC (automatic temperature compensation) correct for temperature, but only once the sample and prism have settled.

Salinity changes slowly, mostly with evaporation and top-off (Module 5). A big jump from yesterday is almost always the measurement, not the tank.

If Then
Bubbles on the conductivity probe The reading is wrong. Tap them off.
Refractometer read right after a room change Temperature not settled. Wait, then read.
Salinity jumped 1.5 ppt since yesterday Almost certainly the measurement. Confirm first.

Check your understanding

A refractometer is carried from an air-conditioned office into a warm fish room and read at once. It shows 32 ppt; yesterday was 35. What do you do?

Let the sample sit on the prism to reach temperature, then read again. If it's still off, check it with 35 ppt fluid before believing it.

Key point Calibrate near 35 ppt, then verify.

11. Probes and controllers

A probe is the part of a meter or controller that sits in the water. It only knows what's around it.

Where a controller probe goes In moving tank water Reads the tank correctly. ! Out of the water Reads the air. The controller heats or cools the wrong way. ! Next to the heater Reads too warm. The heater switches off early. The display only shows what the probe reads, wherever the probe is.

A temperature controller switches a heater or chiller on and off based on what its probe reads. If the probe comes out of the water, it reads the air and the controller heats or cools the tank the wrong way, while its display looks normal. Next to a heater, the probe reads the warm water around the heater, so the heater switches off early and the tank runs cold.

CHECK THE PROBE. Every time you work on a tank, check that controller probes are in the water, in moving water, away from heaters. A controller display is never copied into the log as the tank's reading.

pH probes are stored the way their maker says, usually capped with storage solution. A probe left to dry out can stop reading accurately. If a pH pen reads a little off a reference, its offset is written on a label; apply it yourself before recording (Module 6). ORP probes get dirty and drift, and need cleaning and calibrating too (Module 6).

If Then
Controller probe out of the water It reads the air. The tank is heated or cooled wrongly.
Probe next to the heater Reads too warm. The heater switches off early.
pH pen left uncapped and dry It can stop reading accurately.

Check your understanding

A tank reads 30.5 C on a hand thermometer. The controller display shows 26.5 C and the heater is on. What do you look at first?

The controller probe. If it's out of the water or in the wrong spot, the controller is heating on the wrong number. Put it back in moving tank water and report it.

Key point Probe in moving water, away from heaters. Check it every time.

12. Automated testers

Some systems have machines that test the water by themselves several times a day. They show trends no hand test can, but they're still tools.

Automated testers Many tests a day show the trend Shape only, not to scale. ! Still a tool Reagents run out, tubes clog, calibration slips. Check it against a trusted test.

An automated tester takes water from the system, runs a test with its own reagents, and logs the result. For example, the Neptune Trident tests alkalinity at least four times a day and calcium and magnesium at least twice a day. Its maker states a precision of 0.05 dKH for alkalinity. Testers like this need reagent refills, a container for their waste, and calibration after the reagents are replaced.

Many readings a day show how a number moves through the day and over a week, which a single daily test can't. But an automated tester fails like any tool: reagents run out or go old, tubes clog, calibration slips. Check it against a trusted hand test from time to time, and treat a sudden jump in its readings like any strange number.

If Then
Sudden jump in automated readings Treat it like any strange number. Confirm it.
Reagents just replaced Calibrate before trusting it.

Check your understanding

An automated tester shows alkalinity dropping 1.5 dKH in two hours. Hand tests yesterday were steady. What first?

Treat it as a strange number. Test by hand with a trusted kit. Check the tester's reagents and tubes, and calibrate it if needed.

Key point Still a tool: check it against a trusted test.

13. Lab tests: ICP

Some things can't be measured with hobby kits. For those, a water sample is mailed to a lab.

Hand tests and ICP Hand tests ICP lab test Scope A few parameters Scope Dozens of elements Speed Minutes Speed 1 to 2 weeks Use Daily, weekly numbers Use Traces, contaminants ! Labs can be wrong Results can repeat well and still be off for some elements.

ICP testing (ICP-OES is the common type) measures many elements at once from one water sample. One lab's top package lists 43 parameters, including major elements, trace elements like iodine, strontium and iron, nutrients, and contaminants like copper and lead. The sample is collected in the lab's vials and mailed in. Results take about one to two weeks.

Lab results aren't automatically right. In one independent check, a reef lab tested a certified seawater standard of known values. Its results repeated well, and strontium was within about 1%, but manganese read about five times the true value. The authors concluded such results can show trends and large excesses or shortages, but may not be good enough to fine-tune trace elements.

Use lab tests for what hand tests can't do: trace elements and contaminants, checked now and then. Keep using hand tests for the daily and weekly numbers.

If Then
Need trace elements or copper checked Lab test.
Lab shows one element far off Possible, but labs can be off too. Look at the trend.

Check your understanding

An ICP report shows one trace element far higher than last time. Nothing was dosed. What's a fair way to read it?

It could be real or a lab error. Labs can be off for some elements. Look at the trend across reports, and consider retesting before acting.

Key point Labs for traces and contaminants, hand tests for daily numbers.

14. When a number doesn't make sense

A number that's far from the last one, or physically unlikely, isn't recorded just because a second test agreed. If the tool itself is wrong, it gives the same wrong number twice.

Confirming a strange number ! A strange number Retest carefully Same water, full method Check the tool Standard, or a second device Tool checks out The number is real. Record and report it. ! Tool is off Clean, calibrate, verify, retest.

This is process of elimination. First rule out the reading: test again carefully, with fresh sample and the full method. If it's still strange, rule out the tool: verify it against a standard, or test the same water with a second trusted device. If the tool checks out and the number holds, it's real: record it and follow the form's ranges. If the tool is off, clean it, calibrate it, verify it, and test again.

CONFIRM BEFORE YOU ACT. A number that's far from the last one, or doesn't make physical sense, isn't acted on until it's confirmed. Acting on a wrong number can harm a tank that was fine.

When two different tests disagree on the same water, at least one is wrong. Don't average them. Check each tool against a standard, and trust the one that passes.

If Then
Two tests agree on a strange number The tool may be wrong twice. Verify the tool.
Two tools disagree Don't average. Verify each against a standard.
The tool checks out The number is real. Record it and report it.

Check your understanding

pH reads 9.4 twice in a row. The last test was 8.15. Is two matching tests enough to record it?

No. The same faulty pen gives the same wrong number twice. Check the pen against fresh buffer, or test with a second device, before recording.

Key point Retest, then check the tool. Only then record it.

15. Designing a testing routine

No one can test everything every day. A good routine tests what matters most, most often, the same way each time.

A testing routine Often Fast-changing numbers, and what animals are sensitive to Less often Slow-changing numbers Now and then, by lab Trace elements, contaminants ! After any change Test more: new dosing, new animals, new gear Same tool, same method, same time of day.

Test most often what can change fast and what the animals are most sensitive to. A system with many calcifiers uses alkalinity every day (Module 6), so it needs alkalinity tested more often than a fish-only system. Things that change slowly can be tested less often. Trace elements and contaminants go to a lab now and then.

Test more after any change: a new dosing setting, new animals, new equipment, or a change of salt mix. That's when numbers move.

Keep the method fixed: the same tool, the same method, at about the same time of day. Changing any of those changes the numbers, and then nobody can tell a real change from a different tool. Put calibration and verification on the schedule too.

If Then
A new dosing setting starts Test more often for a while.
Changing test kit brands Test with both for a while to compare.

Check your understanding

A system gets a new two-part dosing setup this week. Should its testing routine stay the same?

No. Test alkalinity and calcium more often while the dosing settles. Numbers move after any change.

Key point Test often what changes fast. Keep the method fixed.

16. Records and trends

A true number in the wrong box is still a wrong number. And one number on its own says much less than a series of them.

Single readings and trends One odd reading Drift ! Odd reading Likely a test error. Retest it. ! Drift Each looks fine alone. Find the cause. Shape only, not to scale.

Log one tank at a time, with that tank's form open. Use the unit and number of decimals the form asks for. Apply any offset label before you enter the number. If you tested at an unusual time, write the time down, so the reading isn't compared as if it were taken at the usual hour. Write what you saw and what you checked, so the next person knows what was already ruled out.

The log turns single readings into trends. One reading far off a steady line is most likely a test error: retest it. A slow creep in one direction, where each reading looks fine on its own, is drift, and only the log shows it (Module 6).

If Then
One reading far off a steady line Likely a test error. Retest.
A slow creep over weeks Drift. Find the cause.
Tested at an unusual time Write the time down.

Check your understanding

You test three tanks in a row, then type all the numbers in at the end. What can go wrong?

Numbers end up on the wrong tank and nobody can tell. Log one tank at a time with its form open.

Key point Right tank, right unit, right decimals, and the time.

17. What you see, and what it means

A big jump from the last reading, nothing else changed Most likely a test error. Retest, then check the tool.
Two tests agree on a strange number The tool may be wrong twice. Verify the tool.
Two tools disagree on the same water Don't average. Verify each against a standard.
A reading of 0.00 Below the detection limit, not none. Report it as low.
Repeated tests agree closely Precise. That doesn't prove accurate.
Fresh standard reads off after calibrating The calibration failed. Recalibrate and verify.
Spots, prints or bubbles on a vial Light is blocked. Clean, dry, retest.
Sample taken at a dosing or top-off outlet Not mixed water. Resample from the main tank.
New kit brand reads differently Kits differ. Compare both on the same water.
Controller display normal, hand reading not Check where the controller probe is.
Automated tester jumps suddenly Confirm with a trusted hand test.
One odd reading on a steady line Likely a test error. Retest.
A small dose measured on the main cup lines Use the rear 1 mL compartment for 5 mL or less.
A slow creep over weeks Drift. Find the cause.

Module 7 test

30 questions. Pass with 27 or more. Opens after you pass every section check in this module. Sign in to start.

Sources

  1. Hanna Instruments: marine phosphate ultra low range Checker HC (HI774) specifications
  2. Bulk Reef Supply: Hanna marine alkalinity Checker HI772 specifications
  3. Hanna Instruments: 9 best practices for using your Hanna Checker
  4. Reefs.com: alkalinity test kit showdown (titration kits compared)
  5. Thermo Fisher: pH calibration procedure for optimal measurement precision
  6. Bulk Reef Supply: how to calibrate and use the Hanna marine salinity tester
  7. Gensou: how to calibrate a refractometer for accurate salinity readings
  8. Bulk Reef Supply: Neptune Systems Trident product spotlight
  9. Bulk Reef Supply: ATI ICP-OES water analysis
  10. Reefs.com: Triton lab ICP-OES testing of a certified artificial saltwater standard
  11. Reef2Reef: loose temp probe, almost complete loss